5.2 Maintenance and servicing
792 Basic IC
149
6 Mounting the pump head
x
Reassemble the components of the pump head
34
34
34
34
as shown
in
Fig. 16
. Tighten the screw
81
81
81
81
by hand. First screw in piston
cartridge
85
85
85
85
manually until the stop is reached and then use
a wrench to turn it through a further
15
°. Firmly tighten the two
valve screw holders
92
92
92
92
with a wrench.
x
Reattach pump head
34
34
34
34
to the pump with the help of the 4
fixing screws
33
33
33
33
. Firmly tighten them with the 6.2621.030 Al-
len key.
To ensure that the pump head is not positioned wrongly, the holes at
the rear for the clamping bolts have different depths, i.e. 1 clamping
bolt is longer than the rest. The deepest hole must naturally accom-
modate the longest bolt. If this is not the case, the pump will not
function properly.
x
Screw connection capillary
35
35
35
35
to pump head
34
34
34
34
(see
Fig. 3
).
x
Connect aspirating tubing to aspirating capillary
32
32
32
32
at the
pump head
34
34
34
34
.
Inlet valve
Outlet valve
6.2824.090 6.2824.080
95
95
95
95
96
96
96
96
97
97
97
97
98
98
98
98
99
99
99
99
100
100
100
100
101
101
101
101
95
95
95
95
96
96
96
96
97
97
97
97
100
100
100
100
99
99
99
99
98
98
98
98
101
101
101
101
Fig. 18: Components of inlet valve
90 and
outlet valve
91
95
95
95
95
Valve housing
96
96
96
96
Sealing ring
(black)
97
97
97
97
Sleeve
98
98
98
98
Sapphire sleeve
The bright side must point towards
the sapphire sphere
99
99
99
99
Sapphire sphere
100
100
100
100
Ceramic holder for sapphire
sphere
The large recess must point in the
direction of the sapphire sphere
101
101
101
101
Seal
The larger opening must point
outwards