4 Operation
792 Basic IC
76
Configuration
The
Configuration
tab in the system settings window contains configura-
tion settings for the 792 Basic IC.
Method
Selection of the method with or without sup-
pressor:
With MSM
method with suppressor
Without MSM
method without suppressor
Suppressor
Suppressor
module:
Autostep with
Automatic switching to the next position each
time the injection valve is switched either to the
Fill
or to the
Inject
position.
Use peristaltic pump
If this option is disabled, the peristaltic pump is
not switched on automatically at the start of a
determination or with
Startup hardware
.
Polarity
Selection of the polarity of the output signal:
+
positive polarity (for anions)
–
negative polarity (for cations)
Hardware settings
Selection of the
Hardware
menu item using the right mouse button
opens the hardware settings window consisting of the two tabs
Hard-
ware
and
Alarm stops
.
Hardware
This tab of the hardware settings window defines general parameters
which are set automatically at power on of the instrument.