2.6 High-pressure pump
792 Basic IC
27
4 Deaerate pump
x
Make sure that the 6.1834.010 aspirating tubing for the high-
pressure pump has been immersed into the eluent.
x
Click the
<On>
button for
IC pump
to switch on the high-
pressure pump.
x
Use the syringe connected to the purge valve
31
to aspirate
air until eluent flows into the syringe.
x
Click the
<Off>
button for
IC pump
to switch off the high-
pressure pump.
x
Close the rotary knob on purge valve
31
31
31
31
by turning it in a
clockwise direction (see
Fig. 3
).
x
Remove the syringe from coupling
24
24
24
24
.
5 Rinse pulsation dampener (if present)
x
Place a beaker beneath the column connection capillary
20
20
20
20
.
x
Click the
<On>
button for
IC pump
to switch on the high-
pressure pump and rinse the pulsation dampener
52
52
52
52
filled
with isopropanol for ca. 10 min with eluent.
x
Click the
<Off>
button for
IC pump
to switch off the high-
pressure pump.
6 Reduce flow rate
x
Reset the original flow rate under
Flow
(e.g.
0.2 mL/min
).
x
Click
<Send to unit>
to send this value to the 792 Basic IC.