![Metasys MULTI SYSTEM 1 Instructions For Use Manual Download Page 8](http://html.mh-extra.com/html/metasys/multi-system-1/multi-system-1_instructions-for-use-manual_1777413008.webp)
EN |
Product description
8
4.
Product description
4.1.
Product description
MULTI SYSTEM TYP 1 is a two-stage, single unit amalgam separator with integrated, static air/water separation and place selection valve.
4.2.
Technical data / performance data
Height [H]
Width [W]
Depth [D]
MULTI SYSTEM TYP 1
Power supply
24 V AC
Frequency
50 / 60 Hz
Max. current consumption
2 A
Max. power consumption
46 VA
Max. ambient temperature
40 °C
Max. liquid fl ow rate
3.0 l/min via the spittoon; 1.5 l/min via the suction side
Negative pressure range
80 mbar - 250 mbar
Separation rate
98.60 %
Collection container volume
300 cm3
Dimensions (H x W x D)
305 x 210 x 104 mm
Dimensions with cover (H x W x D)
340 x 280 x 185 mm
Class
MD class 1
DIBT approval number
Z-64.1-4
Possible suction systems
Wet or dry suction systems
4.3.
Type plate
MULTI SYSTEM TYP 1 Built-in version:
The type plate is located on module 1 and is visible when module 2 is removed (to do this, tilt the locking bracket on
module 2 upwards and remove module 2).
MULTI SYSTEM TYP 1 with cover:
The type plate is located on the cover next to the external display (push the lid of the cover upwards) as well as on
module 1 and is visible when module 2 is removed (see built-in version).