Care and maintenance
| EN
21
4
Clean the probe and probe opening in the fi lter housing.
5
Grease the sealing ring of the probe with multi-silicone grease.
6
Insert the probe into the fi lter housing until you hear it click into place.
7
Carefully push module 2 back into the holder and close the locking bracket.
8
Switch on the main switch.
O
I
7.2.
Maintenance and service
Assembly, modifi cations and repairs may only be carried out exclusively by authorised specialist personnel (see 3.2. Safety
instructions)! For further information and assistance in carrying out repairs, retrofi tting, fault analyses, etc., the METASYS
technical customer service is also available!
Warning:
Switch off the main switch of the treatment unit!
Warning:
Avoid contact with the contents of the amalgam separator!
Warning:
Risk of contamination: To avoid infection, wear personal protective equipment (hand, eye, mouth-nose protection) and
disinfect and clean the device!
7.2.1. Replacement of the collection container
1
Prepare a new collection container and remove the disinfectant bag inside.