SQA-Vision
Service Manual Version 109.13.4
Edition rev.: Jan 2020
22
Release the stopper from the bench at the bottom of the optical column by unscrewing 2
screws using a #2 Allen Key
Line up the false track with the steel rail
Note:
The linear slide
contains many
small bearing balls
that can be
dislodged unless a
false track
is
used to secure
them.
Make sure that the false track is pressed securely against the
steel rail in order to avoid dislodging the small bearing balls
Slide the old camera tray over false track and seal it with the
two plastic stoppers to prevent the tray from slipping out
Remove the front plastic stopper
from the new camera assembly
Slide the entire new camera
assembly back into the optical
assembly over the false track
Re-attach the Camera Stopper
Reassemble the optical assembly by screwing the four screws on the rear section of the
optical column
Reconnect the USB Camera cable connector and put it
back into place
Reattach the base of the optical column on the assembly
Reattach the zoom DC motor, and make sure the cogwheel is matched with the teeth of
the white plastic rack
Reattach the optical column to the front panel
Reinstall the focus knob on the designated axis
Ship to MES both false tracks, the old camera assembly and the empty one in coordination
with the customer support department
Remove the Camera stopper
Lining up the false
track
Slide the old Camera tray
over the false track
Remove the front plastic
stopper
Slide the new camera
assembly back