SQA-Vision
Service Manual Version 109.13.4
Edition rev.: Jan 2020
5
SQA-Vision
SECTION II: System Specifications and Requirements
Dimensions: 32 X 30 X 24 cm
Weight: 7 Kg
AC power supply: 100-240 VAC, 50-60 Hz, 20 VA
Front Panel
Displays: LCD display
Testing: Measurement and Visualization compartments
Other: Multi-button keypad, Focus knob, FOV (Field Of View) knob
Keypad
Operational keys:
I-Button, Service, Enter, Esc, Delete, four cursor buttons and ten
numeric buttons (0-9).
Video control key
:
Zoom In/Out
Measurement Compartment
Sources of radiant energy
- two LEDs for motility and concentration channels
Detector system
- two photo detectors - Motility and Optical Density
Operating System
Analysis Time:
Normal Test – 75 seconds; Low Quality – 2 additional minutes;
Postvasectomy (automated) – 5 minutes
Software:
Resides on flash memory. System can be upgraded from a PC CD-ROM
Motility channel input signal:
Analog, up to 5V.
Concentration channel input signal:
Modulated (kHz) analog, up to 5V.
Rear Panel
Power inlet w/fuse-holder (fuses: 250V,2A), RS232 connector, 2 USB 2.0 B-Type Male,
On/Off switch of the blue surrounding LEDs (located on bottom side of the panel).
Left Side Panel
Power On/Off switch
I-Button port
Visualization Compartment
White LED illumination system with max luminous intensity – 35,000 mcd.
Digital Camera, max resolution: 1280 x 1024 pixels, high resolution of "live" and "frozen"
images provided by capturing a high # of FPS (frames per sec).
Objective: Standard, x40, chromatic aberration correction.
Zoom system for smooth magnification transition between x1188 and x1725
Focus adjustment knob
Field of View Stage knob
SQA-Vision
Maintenance Schedule
Daily:
Clean measurement compartment daily when running samples and after every 10-
15 tests and/or for ANY spillage. Follow manufacturer’s cleaning instructions using
manufacturer cleaning kit (Refer to the appendix section “Cleaning the Capillary
Compartment” in this User Guide).
ONLY use the Manufacturer’s cleaning kit and cleaning brush or damage will
occur to the SQA-Vision film and the system will not operate!
Manufacturer Recommendations
Operate the SQA-Vision away from devices that may cause electronic noise or other
devices causing vibrations such as centrifuges.