Power
issues: fuses,
PSU, power
inlet
Notes:
1.
Turn off the
power supply
and disconnect
the power
supply cable
from the back of
the SQA-Vision
before opening.
2.
Refer to the
Appendix section
for instructions
on how to
address power
supply problems.
Note:
Before checking
the input\ output
connectors of the
main PSU, turn on
the SQA-Vision.
Power issues - fuses, PSU, power inlet.
ISSUE:
The main switch is
ON
, but the power indicator
does not light up and the fan is not working
Check the fuses in the fuses box located on the rear
panel of the SQA-Vision.
Replace the fuse if it is burned out.
Reconnect the power cord and turn the unit back on.
If the unit still does not work, check the input/output
voltage of PSU as instructed below.
How to check the input voltage of PSU:
Check the voltage on the input connector of the main
PSU using a multimeter
The voltage should be ≈220V AC between the
black and the red wires
If no voltage is evident, replace the power inlet (see
the relevant section below)
If voltage is being supplied to the input connector,
check the output connector.
How to check the output voltage of PSU:
Disconnect the output cable from the motherboard.
Check the voltage on the output connector of the main
PSU using a multimeter:
0V: black wire
+5V: red wire
+12V: purple wire
-12V: white wire
If there is no voltage on the output connector or the
voltage is not correct, please replace the main PSU as
instructed below
If the problem persists, please contact Customers
Support.
Replacing the Power Supply Unit –
part #Fpe-E-00146-00
ISSUE
: The power supply and/or cables are not working
Open the SQA-Vision.
Disconnect the input connector of the power supply
unit- location CN1.
Disconnect the output connector of the power supply
unit- location CN2.
Using a #2 Philips screwdriver, unscrew the four
screws that connect the PSU to the rear panel.
Remove the old PSU.
Secure a new PSU to the rear panel using the same
four screws.
Re-connect the input and output cable connectors to
the PSU.
Power Supply
Unit (PSU)
Fuses box
Power Supply Unit
PSU Input connector
Power Supply Output Cable
Unscrew 4 screws of PSU