Carefully pull the optical assembly to gently release it from the front panel
Place the optical assembly on its side
Release the 4 screws holding the Dust Buster using a #2 Philips screwdriver
Release the screw holding the capillary sensor using a #2 Philips screwdriver
Unwind the white plastic cable wrap that holds the capillary sensor and other cables
Remove the damaged capillary sensor from the optical block and unplug the cable
connector from its location on the main board - J13
Attach a new capillary sensor to the optical block using a screw.
Gently re-attach the Dust Buster to the Optical Block using the 4 Phillips screws.
Re-attach the cable to the cable bundle, fasten the cables using the white plastic cable
wrap.
Connect the cable connector to its designated location on the main Board- J13.
Attach the optical column to the front panel of the SQA-Vision using the original 4 screws.
Close the back the front and rear panels of the SQA-Vision using a #2 Philips screwdriver.