92
Operating Instructions - CPC4000
Industrial Pressure Controller
CPC4000
8.6.1 Automatic CPS Installation
The Automatic CPS can be mounted either in a wall/ rack mount configuration by the keyholes (Figure
8.6.1-A) on either or both side of the accessory or placed on a steady surface as a desktop accessory.
The DUT can be mounted either on top of the accessory for convinient test stand setup or on the back
of the accessory for connection with a pressure hose (Figure 8.6.1-B). The not connected port should be
sealed by the plug provided to ensure no leakage.
The pressure connection to the Measure/ Control port of the CPC4000 is established by a pressure tube
with 7/16” SAE adapters. The Automatic CPS is powered by the CPC4000 using a 3 meter long flexible
cable which connects to the Automatic CPS Backplate on the instrument.
[ 38.1mm ]
1.500in
[ 177.8mm ]
7.000in
[ 3.2mm ]
R.125in
[ 6.4mm ]
R.250in
MENSOR
AUTOMATIC CPS LABEL
Figure 8.6.1-A - Side view of Automatic CPS with mounting keyholes
Collection bottle
Device Under
Test (DUT)
Top mounted
DUT port
Connection to the
Automatic CPS Back-
plate of CPC4000
CPC4000 M/C
port connection
Figure 8.6.1-B - Top mounted DUT