Operating Instructions - CPC4000
Industrial Pressure Controller
CPC4000
Figure 10.9-B – Linearization Values
Figure 10.9-B shows some typical values that might be seen in a linearization calibration. In the bottom
right hand corner of this screen is the Graph Icon [
] that, when pressed, reveals a Linearization Er-
ror Graph that gives a visual representation of the errors associated with the values entered in the Linear-
ization screen.
This Linearization error graph shows a scaling that corresponds to the maximum error calculated from
the data entered in the Linearization Matrix. It is a good indication of the overall error of the transducer,
and will quickly reveal any gross data entry errors that have been made. To revert back to the Lineariza-
tion Matrix press the Matrix Icon [ ].
When satisfied that all values have been entered correctly, press the adjust button and then the save but-
ton to save the new calibration data in the transducer memory.
i
Note: After calibration is complete, return to the Calibration Data Application (Section 10.6)
to record the certificate number, calibration interval and the date of calibration. Restoration to
factory calibration can also be completed in this application.