70
Operating Instructions - CPC4000
Industrial Pressure Controller
CPC4000
DHCP
ON or OFF
Reserved for DHCP setup
DHCP?
<sp>(YES or NO)<cr><lf>
Reserved for DHCP setup
DOC
mm/dd/yyyy
Sets the date of cal for the active
transducer.
DOC?
<sp>mm/dd/yyyy<cr><lf>
Returns the date of cal for the active
transducer.
DOM?
<sp>mm/dd/yyyy<cr><lf>
Returns the date of manufacture
Error?
<sp> text description <cr><lf>
Returns the next error in the error
queue
Errorno?
<sp>Enn-text<cr><lf>
Returns pcs400 error code and text
Filter
OFF, LOW, NORMAL, HIGH
Sets the reading filter 0, 80%, 92%,
95%
Filter?
<sp> (filter)<cr><lf>
Returns the reading filter.
Gasdensity
Value in lb/cuft, or “NITROGEN” or
“DRYAIR”
Sets the head pressure gas density in
lb/cuft
Gasdensity?
<sp>+n.nn<cr>
Gets the head pressure gas density in
lb/cuft
Gastemp
Value in degrees F
Sets the head pressure gas tempera-
ture
Gastemp?
<sp>+n.nn<cr>
<lf>
Gets the head pressure gas tempera-
ture
Gateway
nnn.nnn.nnn.nnn
Sets the Ethernet gateway address
Gateway?
<sp>nnn.nnn.nnn.nnn<cr><lf>
Gets the Ethernet gateway address
Height
Value in inches
Sets the head pressure height in
inches
Height?
<sp>+n.nn<cr>
<lf>
Gets the head pressure height in
inches
Highspeed
ON or OFF
Sets the highspeed control mode,
which sets the control behavior, stable
window, stable delay, and rate set-
point
Highspeed?
<sp>(YES or NO)<cr><lf>
Returns the status of the highspeed
mode
Id?
<sp>MENSOR,CPC4000,
ssssss,v.v.vv<cr><lf>
Ssssss is the serial number,v.v.vv is
the CPC 4000 software version.
Instrument_control_
time_on?
<sp>NNNNN<cr><lf>
Returns the amount of time the instru-
ment has been in control mode since
the last reset of the counter, in min-
utes
Instrument_total_con-
trol_time_on?
<sp>NNNNN<cr><lf>
Returns the lifetime amount of time
the instrument has been in control
mode, in minutes
Instrument_time_on? <sp>NNNNN<cr><lf>
Returns the amount of time the instru-
ment has been on since the last reset
of the counter, in minutes