Operating Instructions - CPC4000
19
Industrial Pressure Controller
CPC4000
4.2 Base Instrument
Instrument
Instrument version
Standard: desktop case
Option: 19” rack-mounting kit
Dimensions
See technical drawings
Weight
approx. 127 kg (28 lbs) with all internal options
Warm-up time
approx. 15 min
Display
Screen
7.0” color LCD with resistive touchscreen
Resolution
4 … 6 digits depending on range and units
Connections
Pressure connections
4 ports with 7/16”- 20 F SAE, 1 port with 1/8” F NPT and 1 port with 10-32 UNF female
Filter elements
The instrument has a 40-micron filters on all pressure ports.
Pressure port adapters
Standard: without
Option: 6 mm tube fitting, 1/4” tube fitting, 1/4” female NPT fittings, 1/8” female NPT fittings or 1/8” female
BSP fittings
Barometer port adaptors
Standard: barb fitting
Option: 6mm tube fitting, 1/4“ tube fitting
Permissible pressure media
Dry, clean air or nitrogen (ISO8573-1:2010 Class 5.5.4 or better)
Wetted parts
Aluminum, brass, 316 and 316L stainless steel, Buna N, FKM/FPM, PCTFE, PEEK, PTFE, PPS,
glass-filled epoxy, RTV, ceramic, silicone, silicone grease, Urethane
Overpressure protection
Safety relief valve fixed to reference pressure transducer and adjusted to customized measuring range
Permissible pressure
Supply port
110% FS or 0.69 bar (10 psi), whichever is greater
Measure/Control port
max. 105 % FS
Voltage supply
Power supply
AC 100 ... 240 V, 50/60 Hz
Power consumption
max. 150 VA
Permissible ambient conditions
Storage temperature
-20 ... 70 °C (-4 ... 158 °F)
Humidity
5 … 95 % r. h. (relative humidity, non-condensing)
Compensated temperature range
15 … 45 °C (59 ... 113 °F)
Mounting position
horizontal
Control parameter
Control stability
< 0.005 % FS of the active range
Control mode
precision, high speed and custom
Control time
10 s (regarding a 10 % FS pressure increase in a 50 ml test volume)
Control range
0 ... 100 % FS
Minimum control pressure
0.0017 bar (0.025 psi) over exhaust pressure or 0.05% FS whichever is greater
Overshoots
< 0.3 % FS in high speed control mode (typical <0.1% FS in precision control mode)
Test volume
50 ... 1,000 ccm
Communication
Interface
Standard: Ethernet, IEEE-488, USB, RS-232.
Command sets
Mensor, WIKA SCPI, others optional
Response time
approx. 100 ms
Internal program
up to 24 sequences of 99 steps each