Operating Instructions - CPC4000
Industrial Pressure Controller
CPC4000
Reset_instrument_
control_time_on
Resets the instrument’s time on coun-
ter. See usage application for more
details
Resolution
<n>
Sets the number of significant digits.
(see decpt)
Resolution?
<sp>n<cr><lf>
Returns the number of significant
digits.
(see decpt)
Rfilter
Value in %
Sets the % of the rate filter
Rfilter?
<sp>n.nn<cr>
<lf>
Returns the rate filter.
Rsetpt
Value in current units
Sets the rate setpoint
Rsetpt?
<sp>n.nn<cr>
<lf>
Returns the rate setpoint
Runits
Sec, min, hr
Sets the rate time unit
Runits?
<sp>XXXX<cr><lf>
Returns the rate time unit
Save_cal
Save calibration values
Save_linearity
Save linearity values
Sbaud
9600, 19200, 38400, 57600, 115200
Sets the serial baud rate
Sbaud?
<sp>XXXX<cr><lf>
Returns the serial baud data
Sdata
7 or 8
Sets the serial data bits
Sdata?
<sp>n<cr><lf>
Returns the serial data bits number
Sensor
1,2, P,S, PRIMARY, SECONDARY
Sets the active transducer. Accepts
multiple format types.
Sensor may be specified with P, S, 1
or 2.
Sensor?
<sp>XXXXXXX,1<cr><lf>
Returns active transducer in long
string format
Sensorid?
<sp>Mensor,SN XXXXXX, VN.NN
Returns the active transducer’s serial
number and firmware version
Setpt
value inside upper and lower limits
Sets the control setpoint for the instru-
ment
Setpt?
<sp>XXXXXXX<cr><lf>
Returns the control setpoint in current
units.
Setpt%
Value in % of current range
Sets the control setpoint in % of cur-
rent range
Setptpct
Value in % of current range
Sets the control setpoint in % of cur-
rent range
Setptpct?
<sp>n.nn<cr>
<lf>
Returns the current setpoint in % of
current range
Span
desired pressure or ?
Sets span on active transducer or for
?, clears previous value, must be >
50% FS and has a 1% limit. CALDIS-
ABLE must be OFF/NO.