Operating Manual for Euroklav
®
23V-S
Page 9
3.2 Vacuum
test
________________________________________________________
In order to check the operation of the autoclave in the course of the initial start-up, after long periods without
being used, or after moves, as well as periodically during routine use, a vacuum test should be carried out to
check for leaks.
The vacuum test should be carried out as follows, preferably on the cold autoclave:
Operation
Display message
1. Switch on power, then
after the message
the display will show the
basic parameters
2. Close the door
3
.
Press the "Program"
button several times
until the display shows
"Vacuum test"
4. Press "Start - Stop"
button
5. The
evacuation
pressure has been
reached.
The equalisation period
begins
6. After waiting for the
equalisation (5 min) the
measuring period starts
(running here e.g. for 7
min 52s)
7. After the measuring
period (10 min) the
chamber is ventilated
and
then the leakage rate is
displayed
(if the immediate
printout option is
selected for an attached
printer then a record will
also be printed)
After the message
"Please wait Door
unlocking" this
alternating message will
appear:
Program Start-Stop
V a c u u m t e s t
Program Start-Stop
D o o r
c l o s e d
Program Start-Stop
P l e a s e w a i t
D o o r u n l o c k i n g
Program Start-Stop
1 4 : 2 7 : 1 2
0 . 0 2 b a r 2 5 ° C
Program Start-Stop
L e a k . r a t e 0 . 3 m b a r
Program Start-Stop
V e n t i l a t e
- 0 . 5 6 b a r 2 5 ° C
Program Start-Stop
V a c u t e s t : S t a r t
P r e s s u r e 9 8 0 m b a r
Program Start-Stop
V a c u t e s t 0 0 : 0 0
P r e s s . : 2 4 0 m b a r
Program Start-Stop
V a c u t e s t 0 7 : 5 2
P r e s s . : 2 4 3 m b a r
O p e n d o o r
p l e a s e
Program Start-Stop