![MELAG Euroklav 23V S Operating Manual Download Page 43](http://html1.mh-extra.com/html/melag/euroklav-23v-s/euroklav-23v-s_operating-manual_1766801043.webp)
Operating Manual for Euroklav
®
23V-S
Page 43
8.4
Checking the operation of the autoclave
________________________________
8.4.1
Safety with automatic monitoring ____________________________________________________
The electronic parameter control means that all relevant parameters are constantly monitored a compared
with standard process data, so that error reports can be made immediately. If a program is completed
without problems then on its completion there is an "End" message. The print-out contains a corresponding
report.
The operator of the autoclave can check the progress of the program at any time by means of the values
shown on the display (or after its completion by means of the print-out).
8.4.2 Periodical
bacteriological testing (twice a year) ________________________________________
The German industrial standard DIN 58 946 Part 8 Section 3.2 recommends:
"Periodical testing shall be carried out at the place of installation, e.g. at 6 monthly intervals. They shall
demonstrate that sterilization is carried out satisfactorily when the operating instruction for the small sterilizer
are followed."
Hygiene institutes and regional medical test centres can supply test spores on request and document the
results of sterilization on a test form.
8.4.3 Maintenance
recommendations _____________________________________________________
Regular maintenance of the autoclave is important if it is to have a long life and remain in good working
order.
MELAG recommends that the Euroklav
®
23V-S be serviced annually by a trained technician in accordance
with maintenance instructions for this autoclave. The annual service includes a visual inspection and a test of
operational functions. As well as all essential components and electrical elements, parts are also inspected
for wear and replaced as necessary.
A maintenance reminder appears on the display every two years or after 1000 sterilizations.
Consult your dealer of the MELAG Customer Service if you have any questions relating to servicing and
maintenance.