Operating Manual for Euroklav
®
23V-S
Page 33
6.8 Water
system _______________________________________________________
The Euroklav
®
23V-S also allows the choice between closed-loop and one-way system. In the closed-loop
system the demineralized or distilled water is reused. In the one-way system the demineralized/distilled
water is only used once, which is particularly good for the instruments and the sterilizer, but which leads to
increased water consumption (approx. 500 ml). The water system is selected as follows:
Operation
Display message
1. Hold down "+" button and
also press "-" button.
Select "Function" menu,
submenu "Print"
2. Press "+" or "-" until the
display shows the
submenu "Water system".
3. Press
"Program"
button
and the display shows the
current option, here "One
way“
4. Press the "Program“ to
switch between the option
"One way" and "circle"
(press again to return)
5. When the desired option
has been selected, press
the "Start-Stop“ button to
return to the "Function“
menu and
6. Press the "Start-Stop"
button again to return to
the initial position
6.9 Program
modifications
_______________________________________________
The standard programs are designed to meet most practical operational needs (subatmospheric pulsing,
heating, sterilization, pressure release, drying, and ventilation) and to display the parameters of most interest
(pressure, temperature, time).
The operator is responsible for ensuring that the autoclave is not overloaded , and that the load is arranged
properly to ensure good drying.
There are two standard options "Automatic preheating" and "Additional drying".
Any further program modification to suit specific individual requirements should only be carried out by
authorised personnel, after consultation with your dealer or with the experts at MELAG.
Program Start-Stop
F u n c t i o n :
P r i n t
Program Start-Stop
F u n c t i o n :
W a t e r s y s t e m
Program Start-Stop
W a t e r s y s t e m
O n e w a y
Program Start-Stop
W a t e r s y s t e m
c i r c l e
Program Start-Stop
F u n c t i o n :
W a t e r s y s t e m
Program Start-Stop
1 4 : 2 7 : 1 2
- 0 . 0 2 b a r 2 5 ° C