Warning message
Cause / Remedy
•
This warning appears if the wastewater tank (left tank) is
full.
•
The warning is generated when the water reaches the
maximum level in the container.
•
Empty the tank as follows:
•
Pull on the left plug on the front of the autoclave and
withdraw the emptying pipe as far as possible.
•
Hold the end of the pipe over a container (min. capacity
5 litre) standing on the floor. pull the stopper out of the
pipe and allow the water to drain out.
•
When all the water has drained away, replace the
stopper in the pipe, and push the pipe back into the
opening on the front of the autoclave.
•
The message can then be acknowledged.
Conductivity of the demineralized or distilled water is above
the first limit value, a start is possible by pressing the "Start"
button once more:
For the one-way option
•
Empty water from storage container (right chamber),
clean tank with distilled/demineralized water and refill to
max. with purified water to specifications
•
For the closed-loop option
•
Empty water from both chambers of the internal storage
tank, clean tank with distilled/demineralized water and
refill to max. with purified water to specifications.
•
Water from the MELA
dem
®
47: The demineralization
cartridge in the ion exchanger may be exhausted.
Exchange in accordance with the operating manual.
•
Water from the MELA
dem
®
37: The demineralization unit
may be exhausted. Exchange in accordance with the
operating manual.
•
Water from other purification equipment: Exchange the
demineralization / deionisation unit in accordance with the
manufacturer's instructions.
After taking the appropriate steps, carry out the program start.
When starting for the first time after exchanging the purified
water container, or after maintenance of the water purification
equipment, there may be another report because at first the
supply tube and /or measuring cell will not have been washed
out with fresh, pure water.
Conductivity of the demineralized or distilled water exceeds
the second limit value - a program start is no longer possible:
Proceed as above for "Water quality poor".
W A R N I N G !
N o s t a r t
p o s s i b l e
W a s t e w a t e r t a n k
d i s c h a r g e
A c k n o w l e d g e
w i t h b u t t o n " - "
E x c h a n g e
m o d u l e / c a r t r i d g e
W a t e r q u a l i t y
b a d
A c k n o w l e d g e
w i t h b u t t o n " - "
N o s t a r t
p o s s i b l e
W a t e r q u a l i t y
i n s u f f i c i e n t