Operating Manual for Euroklav
®
23V-S
Page 39
Error report
Cause / remedy
The monitoring time for the ventilation of the pressure
chamber was exceeded.:
•
Check that the pressure releases at the rear of the
chamber are not blocked
•
If this occurs repeatedly, inform your specialist dealer.
Maximum difference between the program duration and the
internal clock exceeded:
If this occurs repeatedly, inform your specialist dealer.
Door not closed properly
Press grip down until contact is made (display should then
show "Door closed")!
If this occurs repeatedly, inform your specialist dealer.
The capillary tube level regulator is open at the start of the
program (error report immediately after start), or the
monitoring time until refilling with demineralised or distilled
water during the program (until the end of sterilization) is
exceeded:
•
This problem can arise because after stopping a program
and immediately restarting - wait for two minutes and try
starting again.
If this occurs repeatedly, inform your specialist dealer.
Maximum permissible time for door locking:
Check the locking bolt can move freely
If this occurs repeatedly, inform your specialist dealer.
The flow monitor for the demineralized / distilled water supply
does not close during the program (see message "Warning no
feed water" - page 34).
The internal testing of the sensors for temperature, pressure
or conductivity showed an excessive deviation, the error can
be reported on switching on the appliance or in the course of a
program:
If this occurs repeatedly, inform your specialist dealer.
The pre-heating has not reached the necessary temperature
within the specified time limit:
If this occurs repeatedly select the option "Automatic
preheating No" and inform your specialist dealer (see Section
6.5).
The maximum preheating temperature was exceeded:
If this occurs repeatedly start the autoclave without
preheating and inform your specialist dealer.
The monitoring time for the pressure release in the outflow
process for the fractionation was exceeded:
•
Check that the flow filter at the bottom of the chamber
close to the door is not blocked
If this occurs repeatedly, inform your specialist dealer.
The limit deviation for internal analog/digital signal conversion
has been exceeded:
If this occurs repeatedly, inform your specialist dealer.
The limit deviation between the two sensors for the steam
temperature has been exceeded:
If this occurs repeatedly, inform your specialist dealer.
M a l f u n c t i o n 1 4 :
W a t e r s u p p l y
M a l f u n c t i o n 1 8 :
S e n s d e f e c t N r .
M a l f u n c t i o n 2 1 :
P r e h e a t i n g
M a l f u n c t i o n 2 2 :
O v e r h e a t i n g
M a l f u n c t i o n 2 6 :
A / D - C o n v e r t i n g
M a l f u n c t i o n 2 7 :
T e m p . s e n s . d e f 1 , 2
M a l f u n c t i o n 4 :
P r e s s u r e r e l e a s e
M a l f u n c t i o n 8 :
T i m e b a s e
M a l f u n c t i o n 9 :
D o o r o p e n
M a l f u n c t i o n 1 2
D o o r l o c k
M a l f u n c t i o n 1 0 :
S t e a m g e n . t o o h o t
M a l f u n c t i o n 2 3 :
C u r r e n t