![Megger PD Scan User Manual Download Page 12](http://html1.mh-extra.com/html/megger/pd-scan/pd-scan_user-manual_1765513012.webp)
Technical description
12
1.2
Technical data
The PD Scan is defined by the following parameters:
Parameter
Value
PD Scan Handheld
Display
3.5" touch display, 320 x 240 pixels
Battery
Li-ion rechargeable battery; 3.6 V, 3.35 Ah
Operating time
>8 hours
Charger input
90 to 264 V AC, 50 / 60 Hz
Charger output
5 V DC
Interfaces
USB type C (charging, headphones, PC)
Bluetooth®
Connection socket for external sensors
Memory
32 MB internal memory /
up to 1000 measurements
IP protection class (in accordance with
IEC 60529 (DIN VDE 0470-1))
IP54
Operating temperature
-10 to 55 °C; 30 °C 93% r.h.
Storage temperature
-20 to 70 °C; 40 °C 95% r.h.
Dimensions W x H x D
220 mm x 80 mm x 30 mm
Weight
420 g
TEV sensors (internal and C900-PD)
Measurement range
Bandwidth
Resolution
Accuracy
Max. number of pulses per cycle
Min. level for pulse counter
-10 to 80 dBmV
2 to 80 MHz
1 dBmV
±1 dBmV
>1226
10 dBmV
Acoustic sensors (internal, AA PR400, AA FR130) and
solid borne vibration sensor ACP 30
Measurement range
Resolution
Accuracy
Converter centre frequency
Laser (AA PR400 only)
-10 to 70 dBµV
1 dB
±1 dB
30, 40 and 80 kHz
Class 2 (max. 1 mW)