12. Reverse L-N Polarity (not available on all instruments)
In countries where the power plug can be reversed in the socket (example: Schuko) the earth leakage tests can be configured to test in
normal “forward” polarity or “normal and reversed” polarity.
Checking the “Reverse L-N Polarity” box in the “Add User”, “Edit User” or “Expert Options” screen enables the forward and reverse
earth leakage testing.
Reversing the test is performed automatically and the worse of the two results are recorded.
The display will show the L
→
N and N
→
L symbols during the leakage tests when this mode is enabled.
12.1
To enable Reverse L-N Polarity testing:
This can only be managed from a “Supervisor” status account. To enable reverse L-N testing on a “User” status account, the change
must be made by a Supervisor.
1)
From the HOME screen use the RIGHT arrow key to select the USER tab and select the Manage Users option.
2)
Press the Edit
Hotkey.
3)
Use the DOWN arrow to select the “Reverse L-N Polarity” check box.
4)
Press the OK key to enable the feature. A tick will appear in the box.
5)
Press the SAVE
Hotkey to store the changes.
Operators who have “Supervisor” status or “Expert” user status may also configure the Reverse L-N Polarity feature in the Expert Options
screen.