background image

will sound on all switch positions except ‘OFF‘ and

.  All selected tests will be

inhibited except for Insulation tests, which will remain available until the voltage
exceeds 55 V. (

BM224

will beep and display a flashing ‘V’. Testing is inhibited).

Default Voltmeter Ranges: 25V - 450V a.c. 50/60 Hz ± 2% ±3 digits

450V - 600V a.c. 50/60 Hz ± 3%
d.c. - unspecified

Automatic Discharge:  Capacitive circuits are automatically discharged
when the test button is released following an insulation test.

Power Supply:

6 x 1,5V cells IEC LR6 type only.

Battery life:

Typically 3000 x 5 second operations (worst 
case, with 

not selected).

Battery cells should not be left in an instrument which may remain unused for
extended periods of time.

Low Battery Indicator:   The low battery indicator symbol 

will appear 

when the battery cells are exhausted.

Battery Replacement:

The rear cover must not be opened if the test

leads are connected. To remove the rear cover, release the screw at the
bottom of the cover and lift the cover upwards.To avoid the possibility of
shock, do not press the test button or touch the fuse when changing
batteries.

Fuse:

500 mA (F) H.B.C.10 kA min (32mm x 6mm)

To check this fuse, select M

Ω, 

open circuit the test leads and press the test

button until a reading is obtained. Display of the fuse symbol 

or an error

code indicates a ruptured fuse. Located behind the rear cover, this fuse can
be replaced by the user. The rear cover must not be opened if test leads are
connected. The replacement fuse must be of the correct type and rating. To
avoid the possibility of shock, disconnect the battery before touching the fuse.

Safety:   The instruments meet the requirements for double insulation to IEC
1010-1 (1995) EN 61010-1 (1995) to Category 

III

*, 300 Volts phase to earth,

440 Volts phase to phase, without the need for separately fused test leads.
*Relates to  transient overvoltage likely to be found in fixed installation wiring.

E.M.C.

In accordance with IEC61326 including amendment 
No. 1

Weight:

530g (including batteries)

Dimensions:

195mm x 98mm x 40mm.

Cleaning:

Wipe disconnected instrument with a clean cloth 
dampened with soapy water or Isopropyl Alcohol 
(IPA).

ACCESSORIES 

Supplied

Part Number

User Guide

6172-231

Synthetic zip-up pouch

6420-132

Test leads, prods & crocodile clips

6220-437

Optional

Fuses (500mA) x 5

6121-289

Fused probe kit FPK8

6111-218

Publications Available

‘A Stitch in Time’

AVTM21-P8B

7

Summary of Contents for BM220 Series

Page 1: ...to become inoperable The Voltage warning does not function if OFF or is selected During an insulation test connections must not be touched After insulation tests capacitive circuits must be allowed to...

Page 2: ...strument On by selecting the range 2 If required zero test lead resistance by firmly shorting both leads together wait for the reading to stabilise and press the test button Display of the symbol conf...

Page 3: ...Voltage 5V 1V Accuracy at 20 C 0 01 9 99 3 2 digits ISC 200mA 10 100 5 2 digits ISC 20mA Lead res comp 0 to 9 99 Hum Rejection 3 error with 1V RMS 0 2 to 50 Continuity Beeper On position the beeper s...

Page 4: ...batteries Fuse 500 mA F H B C 10 kA min 32mm x 6mm To check this fuse select M open circuit the test leads and press the test button until a reading is obtained Display of the fuse symbol or an error...

Page 5: ...roved Repair Companies A number of independent instrument repair companies have been authorised for repair work on most Megger Instruments using genuine Megger spare parts Consult the Appointed Distri...

Page 6: ...038 PO Box 9007 Valley Forge PA 19484 9007 Tel 1 610 676 8500 U S A Fax 1 610 676 8610 Megger SARL 29 All e de Villemomble 93340 Le Raincy Paris Tel 33 1 43 02 37 54 France Fax 33 1 43 02 16 24 This i...

Page 7: ...le confirme la mise z ro Nota La mise z ro des c bles est annul e chaque fois que l instrument est arr t ou que la fonction Arr t automatique est activ e 3 Brancher les sondes d essai sur le circuit i...

Page 8: ...3 c c non sp cifi D charge automatique Les circuits capacitifs sont automatiquement d charg s lorsque le bouton d essai est rel ch apr s un essai d isolement Alimentation lectrique 6 piles 1 5V type I...

Page 9: ...i jedem Ausschalten des Ger ts oder Aktivieren der automatische Abschaltfunktion wieder aufgehoben 3 Pr fspitzen an zu pr fenden Stromkreis anschlie en 4 Auf der Anzeige wird der Widerstand angezeigt...

Page 10: ...C keine Angaben Automatische Entladung Beim Loslassen der Pr ftaste nach einer Isolierpr fung werden die kapazitiven Stromkreise automatisch entladen Stromversorgung sechs 1 5 V Zellen nur IEC LR6 Leb...

Page 11: ...puesta a cero del conductor se cancela cada vez que se desconecta el instrumento o se activa desconexi n autom tica 3 Conecte las sondas de prueba al circuito aislado en prueba 4 El display muestra el...

Page 12: ...apacitivos son descargados autom ticament cuando se suelta el bot n de prueba despu s de una prueba de aislamiento Suministro de energ a 6 pilas de 1 5V de tipo IEC RL6 solamente Vida til de la bater...

Reviews: