All variants of the 

BM220

series of instruments have the following features:

1mA Insulation test current

200mA Continuity test with lead zeroing facility

Combined curved Analogue / Digital display

Continuity Beeper

Auto shut off

Display Back light

Low Battery indication 

In addition, particular models have the following features: 

OPERATION

Preliminary Test lead check 
1.

Before each use of the instrument, visually inspect the test leads, prods
and crocodile clips to confirm that their condition is good, with no damaged
or broken insulation.

2. 

Check continuity of the test leads by firmly shorting the leads together
and read the test lead resistance. measurement directly from the display.

Backlight operation
1.

Turn the instrument backlight ‘On’ by selecting the 

position.

2.

When the backlight is activated, select the desired test position.

3.

On completion, select the ‘OFF’ position to de-activate the backlight and
conserve battery life. 

Continuity Testing
1.

Turn the instrument ‘On’ by selecting the 

range.

2.

If required, zero test lead resistance by firmly shorting both leads together,
wait for the reading to stabilise and press the test button. Display of the

symbol confirms lead zeroing. Note: Lead zeroing cancels each time

the instrument is switched off, or Auto shut off operates.

3.

Connect the test probes to the isolated circuit under test.

4.

The display shows the resistance value. (Maximum 99,9

)

5.

On completion switch to ‘OFF’ position. Alternatively auto shut off operates
after  5  minutes of instrument inactivity.

Continuity Beeper

On the 

position, a continuous beep note

sounds when the test leads make contact with resistance less than 5

Ω. Ι

f

contact is maintained, beeping ceases after a few seconds and the resistance
value is displayed. Resistance greater than 5

is indicated by an intermittent

beep note.

BM221

BM220

BM223

BM224

BM227

* 250 Volt Insulation

* 500 Volt Insulation

* 1000 Volt Insulation

* Default Voltmeter

* Visual Voltage Warning

Leads 
Nulled

Low Battery 

Fuse 

Ruptured

Refer to Safety Warnings before using the instrument

DISPLAY FEATURES 

High

Voltage

1000V

100

200

100 M

300

10 M

400
1 M

V

100k

500

0

1  G

Default
Voltage 

Summary of Contents for BM220 Series

Page 1: ...to become inoperable The Voltage warning does not function if OFF or is selected During an insulation test connections must not be touched After insulation tests capacitive circuits must be allowed to...

Page 2: ...strument On by selecting the range 2 If required zero test lead resistance by firmly shorting both leads together wait for the reading to stabilise and press the test button Display of the symbol conf...

Page 3: ...Voltage 5V 1V Accuracy at 20 C 0 01 9 99 3 2 digits ISC 200mA 10 100 5 2 digits ISC 20mA Lead res comp 0 to 9 99 Hum Rejection 3 error with 1V RMS 0 2 to 50 Continuity Beeper On position the beeper s...

Page 4: ...batteries Fuse 500 mA F H B C 10 kA min 32mm x 6mm To check this fuse select M open circuit the test leads and press the test button until a reading is obtained Display of the fuse symbol or an error...

Page 5: ...roved Repair Companies A number of independent instrument repair companies have been authorised for repair work on most Megger Instruments using genuine Megger spare parts Consult the Appointed Distri...

Page 6: ...038 PO Box 9007 Valley Forge PA 19484 9007 Tel 1 610 676 8500 U S A Fax 1 610 676 8610 Megger SARL 29 All e de Villemomble 93340 Le Raincy Paris Tel 33 1 43 02 37 54 France Fax 33 1 43 02 16 24 This i...

Page 7: ...le confirme la mise z ro Nota La mise z ro des c bles est annul e chaque fois que l instrument est arr t ou que la fonction Arr t automatique est activ e 3 Brancher les sondes d essai sur le circuit i...

Page 8: ...3 c c non sp cifi D charge automatique Les circuits capacitifs sont automatiquement d charg s lorsque le bouton d essai est rel ch apr s un essai d isolement Alimentation lectrique 6 piles 1 5V type I...

Page 9: ...i jedem Ausschalten des Ger ts oder Aktivieren der automatische Abschaltfunktion wieder aufgehoben 3 Pr fspitzen an zu pr fenden Stromkreis anschlie en 4 Auf der Anzeige wird der Widerstand angezeigt...

Page 10: ...C keine Angaben Automatische Entladung Beim Loslassen der Pr ftaste nach einer Isolierpr fung werden die kapazitiven Stromkreise automatisch entladen Stromversorgung sechs 1 5 V Zellen nur IEC LR6 Leb...

Page 11: ...puesta a cero del conductor se cancela cada vez que se desconecta el instrumento o se activa desconexi n autom tica 3 Conecte las sondas de prueba al circuito aislado en prueba 4 El display muestra el...

Page 12: ...apacitivos son descargados autom ticament cuando se suelta el bot n de prueba despu s de una prueba de aislamiento Suministro de energ a 6 pilas de 1 5V de tipo IEC RL6 solamente Vida til de la bater...

Reviews: