Q-Sense Technical Reference Manual
Proprietary
39 of 188
Thermode and Response Unit connectors are color coded.
The Response Unit plug should be connected to yellow receptacle.
The Thermode green and blue plugs should be connected to the green
and blue receptacles respectively.
The SIP/SOPs (signal input and outputs) must be connected to medical
grade safety approved equipment for Canada and the United States by a
recognized certification agency, which is connected to the common
protective earth of the system.
Do not connect any equipment to a Q-SENSE system that has not been
specifically authorized and approved by Medoc.
Figure 10: DC power port
4. Connect the Q-SENSE power supply to an AC outlet and the cable to the DC power port.
4.6
Software Installation
The installation procedure of MEDOC Main Station software can only be performed by a user
with windows system administrator rights. Without appropriate access rights, the
installation procedure may fail.
Users with previously installed versions of Medoc Main Station should review the upgrading
instructions in section
, page
Depending on the current configuration of the computer, MEDOC Main
Station installer may require the installation of additional components.
The installation procedure may take several minutes.
Disable any antivirus that runs on the PC before installing Medoc Main
Station.
The following instructions will guide you through the installation procedure:
1.
Insert the USB drive into an available USB port on the computer
2.
Open the folder to view files.
DC Power Port
Summary of Contents for TSA 2001
Page 1: ...Proprietary 1 of 188 Small Fiber Test Technical Reference Manual ...
Page 99: ...Q Sense Technical Reference Manual Proprietary 99 of 188 Analysis Report Examples ...
Page 109: ...Q Sense Technical Reference Manual Proprietary 109 of 188 ...
Page 181: ...Q Sense Technical Reference Manual Proprietary 181 of 188 Figure 114 Pain Rating Limits ...