Q-Sense Technical Reference Manual
Proprietary
54 of 188
Figure 24: Pre-Test Screen
2.
Click the Start Pre-test button (
).
The system performs a short pre-test, and stabilizes on the Baseline temperature.
3.
After a successful pre-test, the Pre-test button is replaced by a Test Run button (
)
At this point attach the Thermode to the tested subject.
Caution! DO NOT use any electrode paste, jell or other material on the
contact point between the Thermode and the tested subject's skin.
Place the Thermode on the intended stimulation site so that the contact
plate makes the best possible contact with the skin, fastening it lightly
with the attached strap.
It is recommended to first fasten the strap around the area to be
stimulated and then tighten the strap by pulling it an additional 2 cm
before fastening the two ends.
4.
Click the Start button (
).
5.
To pause the stimulation, click the pause button (
).
6.
To stop the stimulation, click the stop test (
) button.
The first click stops the test, and initiates the Finalizing test procedure (accompanied by a
message), which saves relevant information in the Black Box in case of an error.
A second click aborts the test immediately.
7.
Once a test has finished, it is possible to use the mouse curser to display the exact time,
temperature and other available information at any given point.
8.
Click on the Save button
to save the test results and stimulation data.
The system will prompt you to save the stimulation data before starting a
test or exiting the current session.
Summary of Contents for TSA 2001
Page 1: ...Proprietary 1 of 188 Small Fiber Test Technical Reference Manual ...
Page 99: ...Q Sense Technical Reference Manual Proprietary 99 of 188 Analysis Report Examples ...
Page 109: ...Q Sense Technical Reference Manual Proprietary 109 of 188 ...
Page 181: ...Q Sense Technical Reference Manual Proprietary 181 of 188 Figure 114 Pain Rating Limits ...