Q-Sense Technical Reference Manual
Proprietary
187 of 188
1.
TTL Polarity:
1.1.
TTL OUT: Configure the Polarity of TTL OUT pulse: 5V or (-)5V. For 5V: The TTL is 0V
when Idle and 5V when triggered. For (-)5V: The TTL is 5V when Idle and 0V when
triggered.
1.2.
TTL IN: The polarity of the TTL IN does not need to be configured. Q-sense will
automatically detect if it is 5V or (-)5V.
2.
TTL Pulse out Duration:
2.1.
Configure the duration of the TTL OUT. The allowed range is: 50 – 1000 msec.
26.2.3
Standard Event Configuration
TTL trigger shall be sent only if the standard event is configured in the
program
To configure standard events in the required program, follow the steps detailed below:
1.
From the Programs screen, select the program you want to configure standard events for and
double click to open the program editor or click on the edit program button
2.
From the list of parameters, select the Standard Events option
Figure 120: Standard Events in Program Editor
3.
Click the blue arrow on the right to open the standard event editor.
Figure 121: Standard Event Editor
4.
Mark TTL option to activate TTL OUT signal output for various standard events (Onset,
Temperature UP/DOWN, End of duration, End of test).
26.2.4
TTL specifications
Summary of Contents for TSA 2001
Page 1: ...Proprietary 1 of 188 Small Fiber Test Technical Reference Manual ...
Page 99: ...Q Sense Technical Reference Manual Proprietary 99 of 188 Analysis Report Examples ...
Page 109: ...Q Sense Technical Reference Manual Proprietary 109 of 188 ...
Page 181: ...Q Sense Technical Reference Manual Proprietary 181 of 188 Figure 114 Pain Rating Limits ...