Q-Sense Technical Reference Manual
Proprietary
170 of 188
№ Error Message
Solution
29.
Q-sense Thermode was not detected by Q-
SENSE.
The Thermode has been disabled.
1. Make sure that the Thermode is properly
connected to the Q-Sense system.
2. Enable the Thermode through the
Hardware Settings, General menu.
3. Click Yes to try again.
Or
4. Click No to continue working.
The Q-SENSE system could not detect
one of the enabled Thermodes, and
therefore disabled it in order to prevent
any damage.
Follow the displayed instructions to
continue working.
30.
Some of the Safety parameters (Temp Vs.
Time) are violating the Q-SENSE safety
limitations.
1. Click OK to continue working with
current non-approved safety parameters.
2. Click Restore to restore safety
parameters to the Q-SENSE definition.
The Q-SENSE system has detected that
the modified safety limitations do not
comply with safety regulations.
Select one of the presented options and
continue.
In is not allowed to perform any tests on
human subjects when violating safety
limitations without written approval!
31.
Communication between Q-SENSE and PC
lost while in Test Run mode. Error Code:
800C
Communication between the Q-SENSE
system and the PC has been lost.
1.
Restart the Q-SENSE system and
program.
2.
Run Self Test.
3.
Continue working with the system.
If the problem persists, contact Medoc
service.
Summary of Contents for TSA 2001
Page 1: ...Proprietary 1 of 188 Small Fiber Test Technical Reference Manual ...
Page 99: ...Q Sense Technical Reference Manual Proprietary 99 of 188 Analysis Report Examples ...
Page 109: ...Q Sense Technical Reference Manual Proprietary 109 of 188 ...
Page 181: ...Q Sense Technical Reference Manual Proprietary 181 of 188 Figure 114 Pain Rating Limits ...