№ Warning
Explanation
23.
Destination rate is out of range:
Sequence 1 - For a destination
temperature of 23°C, the range must be
[0.1…1].
When performing a Cold EP test there
are restrictions that must be followed:
Temperature change rate must be
within the specified range.
Make sure that your program complies
with these restrictions and continue.
24.
Destination temperature is lower than the
baseline.
Sequence 2 - Return rate must be in the
range [0.1…2]
When performing a Cold EP test there
are restrictions that must be followed:
The temperature return rate must be
within the specified range.
Make sure that your program complies
with these restrictions and continue.
24.1.2
Frequent Error Messages
Table 42: Errors
№ Error Message
Solution
1.
The appearance of this icon, beside a
configurable field or parameter, states
that the parameter value is illegal.
1.
Move the cursor over the sign for
further information.