P3V Veterinary Digital Ultrasonic Diagnostic Imaging System User Manual
- 28
-
Chapter 5 System Control
5.1. Powering On/Off Device
To power on the device
Before powering on this device, check as below:
1.
Check the potential equalization conductor and make sure it is connected properly.
2.
Check all the cables and make sure there is no scrape or crack.
3.
Check the control panel and the monitor and make sure there is no crack.
4.
Check the probe and the connection and make sure there is no scrape or crack.
5.
Check the power socket and the switch and make sure there is no damage.
To power on:
1.
Connect the device to a standard three-pin power supply socket via the power cable, switch
on the AC power switch on the rear panel; Or
Use the battery as the power supply.
2.
Press the power on/off key on the top right control panel, and a startup interface appears.
To shut down the device
1.
Press the
power on/off key on the keyboard and the system displays a confirm dialog box.
2.
Select
Yes
to power off the system.
Or,
If the system breaks down, press the
power on/off key on the keyboard for about six seconds to
shut down the system directly.
NOTE:
Please unplug the AC power cord from the power socket and disconnect the battery if the
device is to remain idle for a long time.
CAUTION
1. You are forbidden to unplug or plug the power cord before switching off the system.
2. Wait approximately five seconds between powering the system off and then on again.
This allows the system to complete its shutdown process.