- 6
-
14.
Periodically have the integrity of the system ground checked by a qualified service
engineer.
15.
To avoid the possibility of electrostatic shock and damage to the system, avoid using
aerosol spray cleansers on the monitor screens.
16.
Do not touch the connector pins on the transducer port.
17.
Parts and accessories used must meet the requirements of the applicable
IEC/EN60601 series safety standards, and/or the system configuration must meet the
requirements of the IEC/EN60601-1.
18.
Do not touch accessible parts of non-medical electrical equipment and the patient
simultaneously.
19.
Any non-medical equipment (such as the external printer) is not allowed to be used
within the patient vicinity (1.5m/6ft.)
20.
Use an extension cord or multi-socket outlet setup to provide power to the ultrasound
system or to the system’s peripheral devices, may compromise the system grounding
and cause the system to exceed leakage current limits.
21.
It is not suggested to use a multiple socket-outlet with the device, if have to, make
sure that the multi-socket complies with the requirement specified in Chapter 16 of
IEC 60601-1:2005, or the multi-socket is with an isolation transformer. And the
multi-socket shall not be placed on the floor.
22.
SHOCK HAZARD - Don't connect electrical equipment, which has not been supplied
as a part of the system, to the multiple portable socket-outlet supplying the system.
23.
SHOCK HAZARD - Don't connect electrical equipment, which has been supplied as a
part of the system, directly to the wall outlet when the non-medical equipment is
intended to be supplied by a multiple portable socket-outlet with an isolation
transformer.
CAUTION
1.
Do not spray cleansers on the system, as this may force cleaning fluid into the
system and damage electronic components. It is also possible for the solvent fumes
to build up and form flammable gases or damage internal components.
2.
Do not use any fluid onto the system surface, as fluid seepage into the electrical
circuitry may cause excessive leakage current or system failure.
3.
To ensure proper grounding and leakage current levels, it is the policy of the
manufacturer to have an authorized representative or an approved third party to
perform all on-board connections of documentation and storage devices to the P3V.
4.
The device and accessories are to be disposed of according to local regulations
after their useful lives. Alternatively, they can be returned to the dealer or the