P3V Veterinary Digital Ultrasonic Diagnostic Imaging System User Manual
- 170
-
Recommended separation distances between portable and mobile
RF communication equipment and the EQUIPMENT or SYSTEM-
For EQUIPMENT or SYSTEM that are not LIFE-SUPPORTING
Recommended separation distances between portable and mobile RF communications
equipment and the P3V
The P3V is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the
P3V can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the P3V as recommended below,
according to the maximum output power of the communications equipment.
Rated
maximum
output
power
of
transmitter
(W)
Separation distance according to frequency of transmitter (m)
150kHz
~
80MHz
d =
P
2
.
1
80MHz
~
800MHz
d =
P
2
.
1
800MHz
~
2.5GHz
d =
P
3
.
2
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output
power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1:
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.