To perform D Trace function (automatic tracing)
1.
Press
Measure
to activate a measurement cursor “+”.
2.
Roll the trackball to highlight
D Trace
and press
Set
.
3.
Select
Auto
and a big “+” is displayed.
4.
Roll the trackball and press
Set
to anchor the start point.
5.
Roll the trackball press
Set
to anchor the end point, the system displays the
results of PS, ED, RI, S/D, PI, TAMAX, etc. in measurement result window.
6.
Roll the trackball and press
Set
to begin a new tracing measurement. You can
measure a maximum of four groups of data. The outcome will be displayed in
the measurement result window.
7.
Press
Measure
to finish and exit.
NOTE
: While automatic tracing, you can add the envelope on the
spectrum upwards or downwards. The higher the sensitivity, the more poor
signal will be enveloped, but also more noises will be enveloped. The
lower the sensitivity, the less the noise signal will be enveloped, but also
some spectrum signal will be lost.
6.6.4. General Report
To preview the general ultrasound report:
Highlight
Print Report
in the B MEAS menu (B Mode Generic Measurement and Calculation
Menu), and press
Set
to display the
general worksheet
dialog box, as shown below:
Figure 6-20 General Worksheet
To edit the general ultrasound report:
Move the trackball to the text box and edit the report, and select
OK
to save the report and close
the dialog box.