P3V Veterinary Digital Ultrasonic Diagnostic Imaging System User Manual
- 105
-
will be displayed. Select TEICHHOLZ or CUBE and press
Set
. Then move the cursor to the
image area and a “+” is displayed.
Move the cursor to the end systolic of left ventricle, and measure LVIDs. The method is similar to
generic M mode distance measurement. LVIDs and ESV will be displayed in the measured result
window.
Move the cursor to the end diastolic of the left ventricle, and then measure LVIDd. The method is
the same as the generic M mode distance measurement method. LVIDd, EDV, SV, EF, and FS
will be displayed in the measured result window.
To enter HR
In the M-cardiac measurement menu, roll the trackball to highlight
Input
. Then select the
secondary menu
Heart Rate
and press
Set
to display a
HR input
dialog box, as shown below.
Figure 9-2 HR Input Dialog Box
Input a suitable value in the HR (bpm) box.
Roll the trackball to highlight
OK
and press
Set
, and after measuring LV, the result of CO will be
displayed in the measured result window.
To enter LVET
In the M-cardiac measurement menu, roll the trackball to highlight
Input
. Then select the
secondary menu
LVET
and press
Set
to display an
ET input
dialog box, as shown below.
Figure 9-3 ET Input Dialog Box
1.
Input a suitable value in the LVET (ms) box.
2.
Roll the trackball to highlight
OK
and press
Set
.