Appendix IV: EMC Information-Guidance and
Manufacture’s Declaration
Guidance and manufacture’s declaration-electromagnetic emissions-
For all EQUIPMENT and SYSTEMS
NOTE:
To protect from EMI, please leave the P3V system away from the EMI sources. For the
technical reasons, electromagnetic immunity is limited to 1 Vrms, otherwise, the
interfaced images may affect the diagnosis and measurements.
Guidance and manufacture’s declaration-electromagnetic emission
The P3V is intended for use in the electromagnetic environment specified below; The customer or the
user of the P3V should assure that it is used in such and environment.
Emission test
Compliance
Electromagnetic environment-guidance
RF emissions
CISPR 11
Group 1
The P3V uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A
The P3V is suitable for use in all establishments, other
than domestic and those directly connected to the public
low-voltage power supply network that supplies building
used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
Complies
Guidance and manufacture’s declaration
–
electromagnetic immunity
–
for all EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration – electromagnetic immunity
The P3V is intended for use in the electromagnetic environment specified below. The customer or the user of P3V should
assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic
environment -guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic tile. If
floor are covered with
synthetic
material,
the
relative humidity should be
at least 30%.