5-8
P11 EXP
Portable Digital Color Doppler Ultrasound System
F4
Figure 5.4:
Biopsy Screen
And one biopsy line displays on the screen, as shown in the following figure.
You can adjust the biopsy offset and biopsy angle if necessary. It the
BIOPSY
is set to
OFF, no submenu displays.
✖
Remarks:
•
The range of the biopsy offset and angle vary with the probe.
•
You should calibrate the biopsy line each time before performing
biopsy. If the position of the biopsy needle is different with that of
the biopsy line, the biopsy cannot be performed.
•
The biopsy offset and angle restore to the factory default settings
after the system is restarted.
5.3.13
Sector Width and Position
User can reduce the imaging area to the region of interest by reducing the sector width and
move the sector position.
SEC.WIDTH
• Press F3 to select SEC.WIDTH.
• turn the function knob to change the sector width.
SEC.POS
• press
to select
SEC.POS
.
• turn the function knob to change the sector position.
BIOPSY OFFSET 12.2
BIOPSY ANGLE 42
Summary of Contents for P11 EXP
Page 1: ...User Manual P11 EXP Ultrasound System Version 1 1 ...
Page 4: ...P11 EXP Portable Digital Color Doppler Ultrasound System 0 2 ...
Page 80: ...P11 EXP Portable Digital Color Doppler Ultrasound System 5 16 ...
Page 102: ...8 8 P11 EXP Portable Digital Color Doppler Ultrasound System ...
Page 118: ...P11 EXP Portable Digital Color Doppler Ultrasound System 10 10 ...
Page 126: ...P11 EXP Portable Digital Color Doppler Ultrasound System 12 6 ...
Page 136: ...P11 EXP Portable Digital Color Doppler Ultrasound System 13 ...
Page 146: ...P11 EXP Portable Digital Color Doppler Ultrasound System A 6 ...
Page 148: ...B 2 P11 EXP Portable Digital Color Doppler Ultrasound System ...