MEATEST, s.r.o.
M-142 Multifunction Calibrator
User Manual v54
89
: EARTh
: VOLTage (?)
<CPD> { ON | OFF | 0 | 1 }
: CURRent (?)
<CPD> { ON | OFF | 0 | 1 }
: AUXiliary (?)
<CPD> { ON | OFF | 0 | 1 }
: ADAPter (?)
: FREQuency
[: CW ] (?)
<DNPD>
: DUTY (?)
<DNPD>
: VOLT (?)
<DNPD>
: ATTE (?)
<DNPD>
: TEMPerature
: UNITs (?)
<CPD> { C | CEL | K }
: SCALe (?)
<CPD> { TS68 | TS90 }
: THERmocouple
[: LEVEl]
[: IMMediate]
[: AMPLitude] (?)
<DNPD>
: RJUNction (?)
<DNPD>
: TYPE (?)
<CPD> { B | E | J | K | N | R | S | T }
: PRT
[: LEVEl]
[: IMMediate]
[: AMPLitude] (?)
<DNPD>
: TYPE (?)
<CPD> { PT385 | PT392 | NI }
: NRESistance (?)
<DNPD>
[SOUR] :FUNC [:SHAP] (?) <CPD> { DC | SIN | PULP | PULS | PULN | RMPA | RMPB |
TRI | LIMS | PWMP | PWMS | PWMN | SQU }
This command sets the shape of the output signal. At the same time, the respective function must be set. E.g. for
:VOLT or :CURR function, FUNC DC, FUNC SIN, or other signal shape must be set. Some functions (:RES,
:CAP) do not require any other settings.
•
DC sets a DC output signal for voltage, current or power modes.
•
SINusoid sets AC output signal for voltage, current or power modes.
•
PULPositive sets AC squarewave signal for voltage or current modes. The amplitude and duty cycle can be
set. The rectangles are positive, i.e. switching between 0 and +amplitude.
•
PULSymmetrical sets AC squarewave signal for voltage or current modes. The amplitude and duty cycle
can be set. The rectangles are symmetrical, i.e. switching between -amplitude and +amplitude.
•
PULNegative sets AC squarewave signal for voltage or current modes. The amplitude and duty cycle can
be set. The rectangles are negative, i.e. switching between –amplitude and 0.
•
RMPA sets AC output signal - ramp shape - for voltage or current modes. The amplitude can be set. -The
output switches between -amplitude a +amplitude.
•
RMPB sets AC output signal - ramp shape - for voltage or current modes. The amplitude can be set. -The
output switches between -amplitude a +amplitude.
•
TRIangel sets AC output signal - triangular shape - for voltage or current modes. The amplitude can be set.
-The output switches between -amplitude a +amplitude.
•
LIMSinusoid sets AC output signal - sinus shape with limitation of amplitude- for voltage or current
modes. The amplitude can be set. -The output switches between -amplitude a +amplitude.
•
PWMPositive sets POS type output signal - digital pulse width modulation.
•
PWMSymmetrical sets SYM type output signal - digital pulse width modulation.
•
PWMNegative sets NEG type output signal - digital pulse width modulation.
•
SQUare sets HSO type output signal - digital pulse output.