by pressing WRITE display button. If you want to skip the calibration point whose calibration
you have already entered, press SKIP display button.
d)
Switch output terminals OFF. Disconnect multimeter and calibrator.
6.
DC current ranges calibration
a)
Connect current input terminals of the multimeter to the +I - -I output terminals of the
calibrator. Select CURRENT DC from the calibration menu.
b)
Select DC current measurement range on external multimeter. Switch output terminals ON
c)
Follow the instructions provided on the calibrator’s display and the DCI table to adjust the
calibrator’s output in the calibration points.
d)
To adjust the calibrator’s output in the calibration points, press SELECT button and use
,
,
,
cursor buttons or potentiometer to adjust the output current. Confirm correctly set value
by pressing WRITE display button. If you want to skip the calibration point whose calibration
you have already entered, press SKIP display button.
e)
Resistance shunt should be used on 2A, 30 A ranges, if standard multimeter does not cover this
range.
7.
AC current ranges calibration
a)
Select CURRENT AC from the calibration menu. Set the same function on external
multimeter.
b)
Follow the instructions provided on the calibrator’s display and the ACI table to adjust the
calibrator’s output in the calibration points.
c)
To adjust the calibrator’s output in the calibration points, press SELECT button and use
,
,
,
cursor buttons or potentiometer to adjust the output current. Confirm correctly set value
by pressing WRITE display button. If you want to skip the calibration point whose calibration
you have already entered, press SKIP display button.
d)
Resistance shunt should be used on 2A, 30 A ranges, if standard multimeter does not cover this
range.
8.
DC power ranges calibration
a)
DC power ranges calibration is performed by calibrating 20 mA, 200 mA, 2 A, 20 A DC
current ranges. There is no need to calibrate voltage ranges.
b)
Select POWER DC from the calibration menu.
c)
Connect a standard ammeter to calibrator’s output terminals +I - -I.
ATTENTION
Calibrator’s Lo and -I terminals are electrically connected.
d)
Follow the instructions provided on the calibrator’s display and the POWER DC table to adjust
the calibrator’s output in the calibration points.