MEATEST, s.r.o.
M-142 Multifunction Calibrator
User Manual v54
55
CURRENT DC function
nominal value [A]
set limits [A]
range [A]
note
0.0
3 n
200 u
zero calibration
0.0
3 n
-200 u
zero calibration
190 u
5 n
200 u
slope calibration
190 u
5 n
-200 u
slope calibration
0.0
20 n
2 m
zero calibration
0.0
20 n
-2 m
zero calibration
1.9 m
50 n
2 m
slope calibration
1.9 m
50 n
-2 m
slope calibration
0.0
100 n
20 m
zero calibration
0.0
100 n
-20 m
zero calibration
19 m
200 n
20 m
slope calibration
-19m
200 n
-20 m
slope calibration
0.0
1 u
200 m
zero calibration
0.0
1 u
-200 m
zero calibration
190 m
2 u
200 m
slope calibration
-190 m
2 u
-200 m
slope calibration
0.0
20 u
2
zero calibration
0.0
20 u
-2
zero calibration
1.9
50 u
2
slope calibration
-1.9
50 u
-2
slope calibration
0.0
300 u
20
zero calibration
0.0
300 u
-20
zero calibration
10
600 u
20
slope calibration
-10
600 u
-20
slope calibration
Tab.8
DC CURRENT
CURRENT AC function
nominal value [A]
set limits [A]
range [A] recommended frequency[Hz]
19 u
5 n
200 u
1000
190 u
50 n
200 u
1000
190u
40 n
2 m
1000
1.9 m
200 n
2 m
1000
1.9 m
200 n
20 m
1000
19 m
2 u
20 m
1000
19 m
2 u
200 m
1000
190 m
20 u
200 m
1000
190 m
20 u
2
500
1.9
200 u
2
500
1.9
1 m
20
120
10
3 m
20
120
Tab.9
AC CURRENT
Other then recommended frequency can be used for calibration. Specification of the calibrator is valid when
recommended frequency is used.