M-142 Multifunction Calibrator
MEATEST, s.r.o.
38
User Manual v54
Multimeter
The calibrator includes a built-in multimeter which can measure many electrical and non-electrical values.
Besides DC voltage and current, it can measure frequency, temperature and when external strain gauge sensors is
connected, even other non-electrical values can be measured. The multimeter can only be connected through
AUXILIARY connector. „Description of controls“ chapter lists the pins of the connector.
Option 40 cable adapter has to be applied to connect the multimeter.
Basic menu
•
Multimeter setting menu can be opened by pressing the METER direct control button. After the button is
pressed, SETUP METER menu opens, which allows to set-up the multimeter. Press EXIT display button to
return to previous menu.
•
Press
or
cursor button to browse the menu options. Active option is always inverted.
•
The parameters of the active option can be changed if the respective function is allowed to be changed. Use
UP, DOWN, CLR, NEXT and display buttons to change the parameters of the active option.
•
Multimeter parameters can also be set using the potentiometer. The potentiometer can be used either to
browse the menu options or to change the value of the active option. Press the knob to switch between these
modes.
•
Keep pressing EXIT display button to return to normal display when you are finished setting up the
parameters.
The manufacturer does not recommend connection of multimeter inputs to
calibrator outputs.
Such connection can result in presence of high voltage at multimeter inputs,
which can damage the multimeter.