M-142 Multifunction Calibrator
MEATEST, s.r.o.
118
User Manual v54
Temperature sensors simulation
temperature scale:
ITS 90, PTS 68
types of sensors:
RTD, TC
A. RTD (resistance) sensors
types:
Pt 1.385, Pt 1.392, Ni
range of R0 setting:
20
to 2 k
temperature range:
-200 to +850
o
C
temperature uncertainty:
0.04
o
C to 0.5
o
C (see table below)
resolution:
0.1
o
Ranges and uncertainties of RTD sensor simulation
type
range –200 – 250
o
C
range 250 – 850
o
C
Pt100
0.2
o
C
0.3
o
C
Pt200
0.1
o
C
0.3
o
C
Pt1000
0.2
o
C
0.4
o
C
Ni100
0.07
o
C
*1
- -
*1
Valid in range –60 to +180
o
C.
Uncertainties in the table are maximal uncertainties of RTD sensor simulation. Actual uncertainty for each set value of
simulated temperature is determined by uncertainty of relevant resistance. Actual temperature uncertainty is displayed on the
calibrator display. Actual uncertainties are always lower than those in above table.
B. TC sensors:
type:
K, N, R, S, B, J, T, E
temperature range:
-250 to +1820
o
C according to the type
temperature uncertainty:
0.4 to 4.3
o
C (see the table below)
resolution:
0.1
o
Ranges and uncertainties of TC sensor simulation
R
range [
o
C]
-50 - 0
0 - 400
400 – 1000
1000 – 1767
uncertainty [
o
C]
1.8 - 1.4
1.4 - 0.7
0.7 - 0.6
0.6 - 0.5
S
range [
o
C]
-50 - 0
0 - 250
250 – 1400
1400 – 1767
uncertainty [
o
C]
1.6 – 1.3
1.3 - 0.8
0.8 - 0.6
0.7 - 0.6
B
range [
o
C]
400 – 800
800 - 1000
1000 – 1500
1500 – 1820
uncertainty [
o
C]
1.7 - 0.9
0.9 - 0.8
0.8 - 0.7
0.7 - 0.6
J
range [
o
C]
-210 - -100
-100 - 150
150 – 700
700 – 1200
uncertainty [
o
C]
0.3 - 0.2
0.2 - 0.1
0.2 - 0.1
0.2
T
range [
o
C]
-200 - -100
-100 - 0
0 – 100
100 – 400
uncertainty [
o
C]
0.4 - 0.3
0.2
0.2
0.1
E
range [
o
C]
-250 - -100
-100 - 280
280 – 600
600 – 1000
uncertainty [
o
C]
0.7 - 0.2
0.2 - 0.1
0.1
0.1
K
range [
o
C]
-200 - -100
-100 - 480
480 – 1000
1000 – 1372
uncertainty [
o
C]
0.5 - 0.2
0.2
0.3 - 0.2
0.3
N
range [
o
C]
-200 - -100
-100 - 0
0 – 580
580 – 1300
uncertainty [
o
C]
0.7 - 0.3
0.3
0.2
0.2
Uncertainties in the table show range of the uncertainties of TC sensor simulation in specified range. Actual uncertainty for
each set value of simulated temperature is determined by uncertainty of relevant DC voltage. Actual temperature uncertainty
is displayed on the calibrator display.