MEATEST, s.r.o.
M-142 Multifunction Calibrator
User Manual v54
29
4.
Setting the power factor (AC power only)
•
If W or Var is indicated, the main power value can be changed by changing the power factor.
Change of power factor does not change the output apparent power.
•
Select P-E mode and then keep pressing the center cursor button until [ _ _ _ _ _ _ _ ] symbols
appear under the power factor symbols (PF = x.xxx LA (LE) or Phase = xxx.x).
•
The value can be set using numeric keyboard and confirmed by pressing LA/LE (°) button or by
pressing ENTER.
•
Main power value is recalculated using new set power factor and existing setting of current and
voltage. The calculation is only made if active or reactive power is displayed.
In the power generation mode, relative deviation cannot be set.
If the power factor is set to define the phase relation of voltage and current, confirmation of entered
value by pressing LA button means positive phase, LE means negative phase.
Setting the energy
Keep pressing the P-E button to switch to the energy generation mode. Auxiliary data display area
shows the time in seconds and the energy delivered to output terminals after pressing the OUTPUT
button given the existing setting of voltage, current, frequency and power factor. Time setting range is
1.1 s to 1999 s.