M-142 Multifunction Calibrator
MEATEST, s.r.o.
54
User Manual v54
VOLTAGE DC function
nominal value [V]
set limits [V]
range [V]
note
0.0
2 u
20 m
zero calibration
0.0
2 u
-20 m
zero calibration
19 m
4 u
20 m
slope calibration
-19 m
4 u
-20 m
slope calibration
0.0
2 u
200 m
zero calibration
0.0
2 u
-200m
zero calibration
190 m
6 u
200 m
slope calibration
-190 m
6 u
-200m
slope calibration
0.0
5 u
2
zero calibration
0.0
5 u
-2
zero calibration
1.9
12 u
2
slope calibration
-1.9
12 u
-2
slope calibration
0.0
20 u
20
zero calibration
0.0
20 u
-20
zero calibration
19
100 u
20
slope calibration
-19
100 u
-20
slope calibration
0.0
200 u
200
zero calibration
0.0
200 u
-200
zero calibration
190
600 u
200
slope calibration
-190
600 u
-200
slope calibration
1000
20 m
1000
slope calibration
-1000
20 m
-1000
slope calibration
Tab.6
DC VOLTAGE
VOLTAGE AC function
nominal value [V]
set limits [V]
range [V] recommended frequency[Hz]
1.9 m
5 u
20 m
1000
19 m
10 u
20 m
1000
19 m
15 u
200 m
1000
190 m
40 u
200 m
1000
190 m
30 u
2
1000
1.9
100 u
2
1000
1.9
200 u
20
1000
19
1 m
20
1000
19
5 m
200
1000
190
10 m
200
1000
190
50 m
1000
1000
750
50 m
1000
500
Tab.7
AC VOLTAGE
Other then recommended frequencies can be used for calibration. Specification of the calibrator is valid when
recommended frequency is used.