MEATEST, s.r.o.
M-142 Multifunction Calibrator
User Manual v54
83
AC current test
Function
Range
Value (A)
Frequency (Hz) Deviation
allowed(A)
A-AC
200.0 uA
0.00019
60
0.305
μ
A-AC
2.0 mA
0.0019
60
1.53
μ
A-AC
20.0 mA
0.019
60
10.5
μ
A-AC
20.0 mA
0.019
120
10.5
μ
A-AC
20.0 mA
0.019
1000
10.5
μ
A-AC
20.0 mA
0.019
10000
109
μ
A-AC
200.0 mA
0.19
60
105
μ
Table VII
AC/DC high current test
Function
Range
Value (A)
Frequency (Hz) Deviation
allowed(%value)
A-DC
2.0 A
1.0
DC
250
μ
A-DC
2.0 A
-1.0
DC
250
μ
A-AC
2.0 A
1.0
60
600
μ
A-DC
30.0 A
10.0
DC
4 m
A-DC
30.0 A
-10.0
DC
4 m
A-AC
30.0 A
10.0
60
16 m
A-DC
30.0 A
30.0
DC
17 m
A-DC
30.0 A
-30.0
DC
17 m
A-AC
30.0 A
30.0
60
45 m
Table VIII
AC/DC power test
Function
Range
Value (VA)
Frequency (Hz) Deviation
allowed(VA)
P 1
480W
480
60
0.336
P 0,5LA
480W
240
60
1.104
P 0,5LE
480W
240
60
1.104
P 1
2400W
2400
60
3.36
P 0,5LA
2400W
1200
60
9.24
P 0,5LE
2400W
1200
60
9.24
Table IX
Voltage = 240V
Resistance test
Function
Range
Value (Ohm)
Frequency (Hz) Deviation
allowed(
Ω
)
O-4W
100.0 Ohm
10.0
DC
6 m
O-4W
100.0 Ohm
100.0
DC
15 m
O-4W
1000 Ohm
1000.0
DC
100 m
O-4W
10k Ohm
10000.0000
DC
1
O-4W
100k Ohm
100000.0000
DC
10
O-4W
1M Ohm
1.6
DC
100
O-4W
10M Ohm
1.7
DC
5 k
O-4W
50M Ohm
5.7
DC
100 k
O-4W
500M Ohm
5.8
DC
2.5 M
Table X
Capacitance test
Function
Range
Value (F)
Frequency (Hz) Deviation
allowed(F)
CAP
1 nF
1.00000e-09
1000
20 p
CAP
10 nF
1.000000e-8
1000
50 p
CAP
100 nF
1.000000e-7
1000
500 p
CAP
1 uF
1.000000e-6
500
10 n
CAP
10 uF
1.000000e-5
120
150 n