M-142 Multifunction Calibrator
MEATEST, s.r.o.
58
User Manual v54
CAPACITANCE function
nominal value [F]
set limits [F]
range [F]
700 p
5 p
700 p – 1 n
1 n
5 p
1 n – 3.3 n
3.3 n
5 p
1 n – 3.3 n
3.3 n
5 p
3.3 n - 10 n
10 n
10 p
3.3 n - 10 n
10 n
10 p
10 n – 33 n
33 n
10 p
10 n - 33 n
33 n
50 p
33 n - 100 n
100 n
50 p
33 n - 100 n
100 n
50 p
100 n - 330 n
330 n
250 p
100 n - 330 n
330 n
250 p
330 n – 1 u
1 u
250 p
330 n – 1 u
1 u
1 n
1 u – 3.3 u
3.3 u
2 n
1 u – 3.3 u
3.3 u
2 n
3.3 u – 10 u
10 u
5 n
3.3 u – 10 u
10 u
5 n
10 u – 100 u
100 u
5 n
10 u – 100 u
Tab.13 Capacitance
FREQUENCY function
amplitude calibration
nominal value [V]
set limits [V]
range [V]
note
0.0
10 u
--
zero calibration
10
1 m
10
slope calibration
Tab.14 Frequency
ANALOG INPUT multimeter function voltage, current, resistance calibration
nominal value [-]
set limits [-]
range [-]
note
0 V
50 u
10 V
zero calibration
10 V
200 u
10 V
slope calibration
0 mA
50 nA
20 mA
zero calibration
19 mA
500 nA
20 mA
slope calibration
0
5 m
200
zero calibration
100
5 m
200
slope calibration
0
50 m
2000
zero calibration
1000
50 m
2 k
slope calibration
0 mV
2 u
20 mV
zero calibration
19 mV
2 u
20 mV
slope calibration
0 mV
7 u
200 mV
zero calibration
190 mV
7 u
200 mV
slope calibration
0 mV
50 u
2 V
zero calibration
1.9 V
50 u
2 V
slope calibration
Tab.15 Multimeter
For calibration of ranges 2, 10 V and 20 mA cable adapter Option 40 must be used. For calibration of resistance
ranges cable adapter Option 60 is necessary. For calibration of voltage ranges 20mV, 200mV, 2V cable adapter
Option 80 has to be applied.