Chapter 8
130
Calibrating the Analog Output Subsystem
This section describes how to configure the DT740 screw terminal panel for an internal or
external meter, and how to use the DT3034 Calibration Utility to calibrate the analog output
subsystems of the board.
Choosing a Calibration Meter
To calibrate the analog output circuitry, you can use either of the following meters:
• The internal A/D converter (ADC) on the DT3034 board.
Using the board’s ADC as an input to the analog output circuitry allows you to calibrate
the analog output circuitry quickly without using external equipment.
• An external precision meter.
The following sections describe how to configure for calibration using either of the supported
meters.
Configuring for the Internal ADC
To calibrate DAC0 using the internal ADC, do the following:
1.
Connect Analog Out 0+ (TB41) to Analog In 2 (TB5).
2.
Connect Analog Out Return to Analog In 2 Return.
To calibrate DAC1 using the internal ADC, do the following:
1.
Connect Analog Out 1+ to Analog In 3.
Signal
DT740 Screw Terminal
Analog Output 0+
TB41
Analog Input 2
TB5
Signal
DT740 Screw Terminal
Analog Output Return
TB42
Analog Input 2 Return
TB6
Signal
DT740 Screw Terminal
Analog Output 1+
TB43
Analog Input 3
TB7
Summary of Contents for Data Translation DT3034
Page 2: ...DT3034 User s UM 22359 M Manual Title Page...
Page 5: ......
Page 11: ...Contents 10...
Page 15: ...About this Manual 14...
Page 16: ...15 1 Overview Features 16 Supported Software 18 Accessories 19 Getting Started Procedure 20...
Page 22: ...Part 1 Getting Started...
Page 23: ......
Page 33: ...Chapter 2 32...
Page 39: ...Chapter 3 38...
Page 68: ...Part 2 Using Your Board...
Page 69: ......
Page 111: ...Chapter 6 110...
Page 123: ...Chapter 7 122...
Page 135: ...Chapter 8 134...
Page 141: ...Chapter 9 140...
Page 153: ...Appendix A 152...
Page 154: ...153 B Connector Pin Assignments...