Chapter 8
126
Follow the instructions on
.
Configuring for an External Reference
To calibrate the analog input circuitry using an ex9.3750 V reference, do the following:
1.
Connect Analog In 0 to the positive side of the precision voltage source.
2.
Connect Analog In 0 Return to the negative side of the precision voltage source.
3.
Connect Analog In 0 Return to Analog Ground.
4.
Connect Analog In 1 and Analog In 1 Return to Analog Ground.
Signal
DT740 Screw Terminal
Analog Input 1
TB2
Analog Input 1 Return
TB4
Analog Ground
TB34
Signal
DT740 Screw Terminal
Analog Input 0
TB1
Signal
DT740 Screw Terminal
Analog Input 0 Return
TB2
Signal
DT740 Screw Terminal
Analog Input 0 Return
TB2
Analog Ground
TB34
Signal
DT740 Screw Terminal
Analog Input 1
TB3
Analog Input 1 Return
TB4
Analog Ground
TB34