Chapter 4 Operation
MX-2100/2104
Installation and Operation Manual
4-14
Operating Instructions
After the self-test, the MX-2100 prompt should already be displayed on the
terminal screen after the ON-LINE announcement. If you see
PASSWORD>
this indicates that password protection is enabled.
2. In this case, type:
'password'
and press ENTER
where 'password' stands for the current password (four to eight characters).
The default password is
RAD
.
If your password is accepted, you will see the prompt KM2100>.
MX-2100 is now in session under your control.
On your terminal, you will see the prompt:
KM2100>
or
KM2104>
On the MX-2100 front panel, you will see the message:
TERMINAL ON LINE
The front panel controls are disabled as long as MX-2100 is under remote control.
3. Before you continue, use the F command to select the proper cursor control
codes, otherwise the DEF commands may not work properly.
While the supervision terminal is in session with MX-2100, the MX-2100 local
operator can regain control by disconnecting the cable from the MX-2100 SP
connector, or by sending the EXIT command. MX-2100 will automatically return to
front panel control if no commands are received for a certain time (controlled by
the LOG_OFF parameter). This time-out can be disabled.
Starting a Session - Multiple MX-2100
When one terminal is used to control several MX-2100 units connected via
modems, node addresses are assigned to each MX-2100. The node addresses, in
the range of 1 through 255, are assigned during the first session, by means of the
command NODE.
If you are using a multidrop configuration, do not assign address 0 to any MX-2100
connected to this terminal.
Note
Note