4-10
Section
4.6
Programming Deployment Parameters
After the sample bottles are filled and attached to the rotator, the next step for a deployment is
programming the Deployment Parameters.
Instrument Current Consumption
The Sediment Trap holds user-replaceable, drop-in “C” cell alkaline batteries with a total capacity
of 10,000 mAh. Battery capacity for a planned deployment can be estimated using the Instrument
Current Consumption tables in this section..
Instrument
Current
Consumption
Controller electronics ON
9.0 mA
Controller electronics LOW POWER SLEEP
0.22 mA
Bottle Rotation Consumption
21 Bottles
26 sec/rotation, 380 mA
13 Bottles
13 sec/rotation, 380 mA
System Deployment / Event Parameters for this example
Number of Events
21 or 13 samples
Length of Deployment
1 Year