6-3
Section
6.2
Upon Returning to your Lab
Several maintenance steps should be completed when the Sediment Trap returns from a deployment
before longer term storage.
Upon Returning to your Lab
Step
Note
1
Rinse and scrub the cone, housing, motor and
rotator with fresh water.
Take extra care around the flow
straightener.
2
Check Oil Bladder in Drive Motor Assembly
Add silicone oil only if the bladder is
low (see ‘Check the Oil Bladder in the
Drive Motor Assembly’ in this
chapter).
3
Remove and inspect controller housing internals
(check o-rings, remove batteries)
See ‘Remove and Inspect Controller
Housing Internals’ in this chapter.
4
Remove rotator and rinse ball bearings with fresh
water
See ‘Remove Rotator and Rinse Ball
Bearings’ in this chapter.
5
Reinstall the components onto the frame
See ‘Reinstall Components onto the
Frame’ in this chapter.
6
Cover top/funnel with clean plastic for storage
Take care not to put weight on the
honeycomb baffle