Appendix B-10
Offload/Display Data – EEPROM Pressure and/or Temperature Sensor
Selecting <4> displays the backup EEPROM. Pressure and/or temperature data is also stored in
EEPROM.
Figure B-12: EEPROM Data Including Pressure and Temperature
Selection [1] ? 4
During deployments, a backup copy of the instrument data
is written to non-volatile EEPROM storage. This allows
for data recovery in the event the instrument data
is no longer resident in active memory.
The sensor time history is recorded in the disk
file "DATA.DAT" which will be displayed here.
(It can also be copied directly from the disk.)
To copy the EEPROM cache file to a disk file:
Start the capture file, now.
Then, press any key to start the transfer.
The cache file remains in EEPROM until overwritten during the
next deployment.
Event Summary:
Event 01
Scheduled start time: 04/10/17 21:00:01
Event start time: 04/10/17 21:00:01
Event stop time: 04/10/17 21:00:30
Event duration: 29 seconds
Aligned Battery Temperature Tilt Heading
Start: Y 17.8 -18.9 °C 92° 271°
Stop: Y 16.5 -18.8 °C 92° 271°
Event 02
Scheduled start time: 04/10/17 21:30:01
Event start time: 04/10/17 21:30:01
Event stop time: 04/10/17 21:30:30
Event duration: 29 seconds
…
Display shortened to save space
Summary of Contents for Mark 78H-21
Page 1: ...Sediment Trap User Manual...
Page 6: ...TOC 4...
Page 28: ...1 18 Notes...
Page 44: ...3 8 Notes...
Page 74: ...5 8 Notes...
Page 104: ...A 4 Notes...
Page 128: ...A 4 Notes...
Page 151: ...Appendix D 3 Test Cable Wiring Diagram Figure D 3 Test Cable Wiring Diagram...
Page 160: ...Appendix E 8 Notes...
Page 168: ...Appendix E 8 Notes...